Design of experiments in medical physics: Application to the AAA beam model validation

Abstract :

Purpose

The purpose of this study is to evaluate the usefulness of the design of experiments in the analysis of multiparametric problems related to the quality assurance in radiotherapy. The main motivation is to use this statistical method to optimize the quality assurance processes in the validation of beam models.

Method

Considering the Varian Eclipse system, eight parameters with several levels were selected: energy, MLC, depth, X, Y1 and Y2 jaw dimensions, wedge and wedge jaw. A Taguchi table was used to define 72 validation tests. Measurements were conducted in water using a CC04 on a TrueBeam STx, a TrueBeam Tx, a Trilogy and a 2300IX accelerator matched by the vendor. Dose was computed using the AAA algorithm. The same raw data was used for all accelerators during the beam modelling.

Results

The mean difference between computed and measured doses was 0.1 ± 0.5% for all beams and all accelerators with a maximum difference of 2.4% (under the 3% tolerance level). For all beams, the measured doses were within 0.6% for all accelerators. The energy was found to be an influencing parameter but the deviations observed were smaller than 1% and not considered clinically significant.

Conclusion

Designs of experiment can help define the optimal measurement set to validate a beam model. The proposed method can be used to identify the prognostic factors of dose accuracy. The beam models were validated for the 4 accelerators which were found dosimetrically equivalent even though the accelerator characteristics differ.

Type de document :
Article dans une revue
Liste complète des métadonnées

https://hal.univ-angers.fr/hal-02525527
Contributeur : Okina Université d'Angers <>
Soumis le : mardi 31 mars 2020 - 03:07:16
Dernière modification le : mercredi 10 juin 2020 - 12:16:02

Identifiants

Collections

Citation

S. Dufreneix, Christophe Legrand, C. Di Bartolo, M. Bremaud, J. Mesgouez, et al.. Design of experiments in medical physics: Application to the AAA beam model validation. Physica Medica, Elsevier, 2017, 41, pp.26-32. ⟨10.1016/j.ejmp.2017.05.068⟩. ⟨hal-02525527⟩

Partager

Métriques

Consultations de la notice

20